17 Feb 2011 - ACIS peaks and sigmas
24 Mar 2011 - ACIS line fluxes
04 May 2011 - XIS peaks versus COR
04 May 2011 - IACHEC plots: ACIS & XIS peaks, FWHMs & line fluxes
21 Jun 2011 - Is line centroid a good proxy for CTI?
17 Oct 2011 - Figures for paper?
31 Oct 2011 - Trailing charge
31 Oct 2011 - Figures for paper
21 Nov 2011 - OLD Schematic figures for paper
02 Dec 2011 - CTI change at -120C and -90C on ACIS
20 Dec 2011 - Schematic figures for paper
02 Feb 2012 - XIS Perseus analysis for non-CTI changes
09 Feb 2012 - XIS Perseus analysis for non-CTI changes, part II
05 Sep 2012 - ACIS fits with Gehrels weighting
05 Sep 2012 - cal source plots
14 Sep 2012 - XIS fits with Gehrels weighting
14 Sep 2012 - XIS background trend from night Earth data
24 Sep 2012 - new cal source plots
25 Sep 2012 - perseus intercepts
31 Oct 2012 - XIS BG trend from night Earth data, updated
31 Oct 2012 - XIS + ACIS background spectra with lines ID'd
17 Jan 2014 - XIS + ACIS calibration source spectra overplotted
17 Jan 2014 - XIS + ACIS background spectra with lines ID'd, version 2
13 Mar 2014 - new cal source plots

SPENVIS work has its own page: spenvis

...

Anchor
anchor1
anchor1
17 Feb 2011 - Catherine's attempt to duplicate Bev's monthly XIS plots for ACIS I3 (FI) and S3 (BI). Each data point is a fit to 0.1 years worth of cold ACIS calibration source data, filtered to include only the upper corner regions like XIS and grades G02346. After some testing as to which fitting procedures were less sensitive to background variations (important for ACIS probably not as much for XIS), and robust against low signal-to-noise data, I chose to use a Gaussian plus a linear background component (mpfitfun.pro with some limits to prevent non-physical fits).

...

ACIS has a known slow change in the gain as a function of time as measured very close to the framestore where CTI should be negligible. For all of the CCDs except I0 and I2 it is monotonically decreasing at a rate of ~1 ADU/yr at 5.9 keV. (The gain change on I0 and I2 is pathological with jumps and annual trends that aren't relevant, so I don't use them here.) See http://space.mit.edu.ezproxyberklee.flo.org/~cgrant/gainImage Removed and http://space.mit.edu.ezproxyberklee.flo.org/~cgrant/lineImage Removed for example plots.

...

(I am also ignoring the issue of changing serial CTI. For ACIS this is negligible, but might not be for Suzaku?)

...

Anchor
anchor6
anchor6
Oct 17, 2011 - Catherine

Paper appropriate figures?

I've attached the encapsulated postscript files for the figures from my HEAD2011 poster. Not sure if they follow the A&A rules but they have consistent notation with nice thick lines. They use color, but work OK in B/W as well.

ACIS peak vs time
ACIS FWHM vs time
XIS peak vs time
XIS FWHM vs time
XIS peak vs COR
XIS FWHM vs COR

...

Anchor
anchor7
anchor7
Oct 31, 2011 - Catherine

ACIS trailing charge plots, to be compared to the XIS equivalent here (http://space.mit.edu.ezproxyberklee.flo.org/XIS/monitor/ccdperf/)

Image Added

...

Anchor
anchor8
anchor8
Oct 31, 2011 - Catherine

Figures for paper (new versions)

ACIS peak vs time
ACIS FWHM vs time
ACIS bkg vs time
XIS peak vs time
XIS FWHM vs time
XIS peak vs COR
XIS FWHM vs COR

...

Anchor
anchor9
anchor9
Nov 21, 2011 - Eric

Schematic figures for paper (removed)

...

Anchor
anchor10
anchor10
Dec 2, 2011 - Catherine

CTI change at -120 and -90C on ACIS

At my review, Mark and I talked about the problem of separating the different radiation environment from the different focal plane temperatures. I remembered that we had taken data with ACIS at -90C at two different times, in Sept 1999 and Aug 2005, so we can at least see how different the CTI evolution is on ACIS at both temperatures.

Image Added
ACIS peak vs temperature for S2 and S3

The top panel is for ACIS-S2, the bottom for ACIS-S3. (There's isn't any data for I3 in 1999, so I've switch so S2 which is also an FI device but has higher CTI.) The data points are showing the centroid pulseheight as a function of focal plane temperature (measured in the same regions and the same way as everything above.) The later data has lower pulseheights which is consistent with the CTI increase. For both devices the size of the increase is larger at -90C than it is at -120C.

More specifically...

  • S3 at -119C, the peak drops by 0.8%
  • S3 at -90C, the peak drops by 2.1%
  • S2 at -119C, the peak drops by 2.0%
  • S2 at -90C, the peak drops by 9.9%

For S3, the drop is 2.7 times faster at -90C than at -120C. For S2, the drop is 5.1 times faster at -90C than at -120C. To compare to XIS, the FI CCDs on XIS drop 5 times faster than ACIS, the BI CCDs drop 9 times faster than ACIS. (The XIS numbers are with charge injection on, but before the BI charge injection was increased.) So the difference in CTI evolution between ACIS and XIS may be entirely due to the focal plane temperature?

(The FWHM of ACIS-S2 at -90C is so large that the Mn-K line is hopelessly entangled with the Mn-Kbeta and Ti-K lines, so I haven't extended this analysis to FWHM.)

...

Anchor
anchor11
anchor11
Dec 20, 2011 - Eric (updated from Dec 06, 2011)

Schematic figures for paper (v3)

Image Added
ACIS schematic PS

Image Added
XIS schematic PS

...

Anchor
anchor12
anchor12
Feb 02, 2012 - Eric

Analysis of XIS3 Perseus data from 0<ACTY<128, to see how well we might constrain non-CTI gain changes. Line centroid is fit to 0.1%, might have issues from differing kT. These data have been gain- and CTI-corrected, so this is just a proof of concept.

Image Added

Image Added
XIS3 Perseus spectrum PDF

Image Added
Perseus model spectra with kT = 4 keV and 7 keV

...

Anchor
anchor13
anchor13
Feb 09, 2012 - Eric

Analysis of XIS1,3 Perseus data, re-PHA'd and re-graded by Bev. GRADE_BEV=0,2,3,4,6 were used along with PHA_BEV, which have no CTI or gain correction. A powerlaw+Gaussian was fit to the Fe XXV K line complex and best-fit center is plotted for segments 1 and 2 (=B and C), and 128-row bins at the bottom, middle, and top of the CCD.

Image Added
XIS1 Perseus line center trend PDF
XIS1 Perseus fit spectra PDF

Image Added
XIS3 Perseus line center trend PDF
XIS3 Perseus fit spectra PDF

...

Anchor
anchor14
anchor14
Sep 05, 2012 - Catherine

Figures 6 and 9 (ACIS peaks and FWHMs vs time) using Gehrels weighting in the fitting and showing the error bars. In general, the errors bars are very small. Gehrels weighting doesn't change the results much.

Image Added

Image Added

...

Anchor
anchor15
anchor15
Sep 05, 2012 - Bev

calibration source plots. I fit the peak in adu for each quad separately, then re-processed using the main fe55 peak as the gain, then plotted quads A&D combined. I normalized by dividing by the total number of counts in the spectrum from 0 to 10 keV but only plotted 1 to 10. g02346 events only.

both side by side
both stacked
acis only
xis only

...

Anchor
anchor16
anchor16
Sep 14, 2012 - Catherine

Figures 5 (xis peaks vs time), 8 (xis fwhm vs time), 12 (xis peak vs cor) and 13 (xis fwhm vs for) using Gehrels weighting in the fits and showing the error bars. The Gehrels results are similar to what's in the paper except for fwhm vs time which increasing more slowly in the Gehrels fits.

Image Added
Image Added
Image Added
Image Added

...

Anchor
anchor17
anchor17
Sep 14, 2012 - Eric

Trend of the XIS NXB from night Earth data, extracting a 512x512 box from the chip center, using only COR2>6 data with normal SCI on (XIS1 at 2 keV level). XIS1 has a clear downward slope of -0.003 cts/s/yr +/- 20%, or about 3% per year. XIS0,3 are consistent with no change. Perhaps this is actually a gain issue with XIS1.

Look at the PDF also.

Image Added

Update post-meeting: Now including all COR2, 5-15 keV (to reduce gain effects around 7 keV). XIS1 still changes by 4% +/- 1% per year, though with lots of scatter. +/-1-sigma limits for FIs: +0%/-1% (XIS0), +0.5%/-0.5% (XIS3).

Look at the PDF also.

Image Added

XIS NXB spectra from Tawa et al 2008 (Fig 1):

Image Added

...

Anchor
anchor18
anchor18
Sep 24, 2012 - Bev

New version of cal source plots.

Image Added

eps version for .tex.

...

Anchor
anchor19
anchor19
Sep 25, 2012 - Bev

I took the intercept of peak vs row for the main iron line in the perseus data
and fit it versus time. The summary is that the intercept is dropping less than half a percent per year for the higher CI levels and less than 1 percent per year for the original, lower, BI level.

CCD

quad

CI

 

adu/year

error (edu/year)

XIS1

c0

higher CI

3 SRs

-5.39

21.81

XIS1

c0

original CI

12 SRs

-15.24

1.56

XIS1

c1

higher CI

3 SRs

1.63

3.59

XIS1

c1

original CI

12 SRs

-8.94

0.47

XIS1

c2

higher CI

3 SRs

-2.42

4.03

XIS1

c2

original CI

12 SRs

-11.25

0.48

XIS1

c3

higher CI

3 SRs

6.52

10.35

XIS1

c3

original CI

12 SRs

-11.54

1.59

XIS3

c0

FI CI

13 SRs

-5.41

0.95

XIS3

c1

FI CI

13 SRs

-3.66

0.32

XIS3

c2

FI CI

13 SRs

-3.54

0.26

XIS3

c3

FI CI

12 SRs

-5.07

0.82

ps of plots. individual eps files available on request.

...

Anchor
anchor20
anchor20
Oct 31, 2012 - Eric

Trend of the XIS NXB from night Earth data, extracting all but the cal source regions and bad regions of each chip, using all COR2 data with normal SCI on (XIS1 at 2 keV level). See previous post.

Here it is in 4 week bins, with the XIS1 y errorbars corresponding to the counting statistics. First using FTOOL lcurve:

Image Added

Second doing this by hand in xselect by and-ing a 4 week time range with the GTIs:

Image Added

I don't understand the scatter; each bin contains about 50 ksec of data, 1000's of counts, and should span a similar distribution of COR2, yet there is an intrinsic ~ 25% scatter, far higher than the < 5% quoted by Tawa. Restricting COR2>=10 does not change this, nor do larger time bins. Perhaps something is wrong with the GTIs and thus the effective exposure time calculation.

...

Anchor
anchor21
anchor21
Oct 31, 2012 - Eric

Background spectra for XIS and ACIS with lines ID's a lo Tawa:

Image Added

Also the EPS for the paper.

...

Anchor
anchor22
anchor22
Jan 17, 2014 - Bev

Calibration source spectra for XIS and ACIS:

Image Added

Also the EPS for the paper.

...

Anchor
anchor23
anchor23
Jan 17, 2014 - Eric

Background spectra for XIS and ACIS with lines ID's a la Tawa,
now with "FI CCD" and "BI CCD" identified.

Image Added

Also the EPS for the paper.

...

Anchor
anchor24
anchor24
Mar 12, 2014 - Bev

Calibration source spectra for XIS and ACIS:

Image Added

Also the EPS for the paper.

...