Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Contents:
17 Feb 2011 - ACIS peaks and sigmas
24 Mar 2011 - ACIS line fluxes
04 May 2011 - XIS peaks versus COR
04 May 2011 - IACHEC plots: ACIS & XIS peaks, FWHMs & line fluxes
21 Jun 2011 - Is line centroid a good proxy for CTI?
17 Oct 2011 - Figures for paper?
31 Oct 2011 - Trailing charge
31 Oct 2011 - Figures for paper
21 Nov 2011 - OLD Schematic figures for paper
02 Dec 2011 - CTI change at -120C and -90C on ACIS
06 Dec 2011 - Schematic figures for paper

...

Anchor
anchor1
anchor1
17 Feb 2011 - Catherine's attempt to duplicate Bev's monthly XIS plots for ACIS I3 (FI) and S3 (BI). Each data point is a fit to 0.1 years worth of cold ACIS calibration source data, filtered to include only the upper corner regions like XIS and grades G02346. After some testing as to which fitting procedures were less sensitive to background variations (important for ACIS probably not as much for XIS), and robust against low signal-to-noise data, I chose to use a Gaussian plus a linear background component (mpfitfun.pro with some limits to prevent non-physical fits).

...

Schematic figures for paper

Image Removed
ACIS schematic PS

Image Removed
XIS schematic PS

(removed)

...

Anchor
anchor10
anchor10
Dec 2, 2011 - Catherine

...

(The FWHM of ACIS-S2 at -90C is so large that the Mn-K line is hopelessly entangled with the Mn-Kbeta and Ti-K lines, so I haven't extended this analysis to FWHM.)

...

Anchor
anchor11
anchor11
Dec 06, 2011 - Eric

Schematic figures for paper (v2)

Image Added
ACIS schematic PS

Image Added
XIS schematic PS

...

Attachments

...