17 Feb 2011 - Catherine's attempt to duplicate Bev's monthly XIS plots for ACIS I3 (FI) and S3 (BI). Each data point is a fit to 0.1 years worth of cold ACIS calibration source data, filtered to include only the upper corner regions like XIS and grades G02346. After some testing as to which fitting procedures were less sensitive to background variations (important for ACIS probably not as much for XIS), and robust against low signal-to-noise data, I chose to use a Gaussian plus a linear background component (mpfitfun.pro with some limits to prevent non-physical fits).
...
24 Mar 2011 - The line normalizations that come out of the above fits. The first is a linear plot of the normalization versus time showing the obvious signature of radioactive decay.
Then comparing the data to the expected half-life 2.737 years for Fe-55.
...
Attachments |
---|